CVE-2024-48966: Baxter LIFE2000 Ventilation System

Critical severity, CVSS 10.0. EPSS: 0.7% chance of exploitation in the next 30 days.

The software tools used by service personnel to test & calibrate the ventilator do not support user authentication. An attacker with access to the Service PC where the tools are installed could obtain diagnostic information through the test tool or manipulate the ventilator's settings and embedded software via the calibration tool, without having to authenticate to either tool. This could result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.

Affected products

  • Baxter LIFE2000 Ventilation System: up to and including 06.08.00.00
  • Baxter LIFE2000 Ventilator Firmware: before 6.08.00.00 (fixed in 6.08.00.00)

Published 2024-11-14. Last modified 2026-06-17.