CVE-2021-33886: Bbraun SPACECOM2
High severity, CVSS 8.8. EPSS: 0.8% chance of exploitation in the next 30 days.
An improper sanitization of input vulnerability in B. Braun SpaceCom2 prior to 012U000062 allows a remote unauthenticated attacker to gain user-level command-line access by passing a raw external string straight through to printf statements. The attacker is required to be on the same network as the device.
Affected products
- Bbraun SPACECOM2: before 012u000062 (fixed in 012u000062)
Published 2021-08-25. Last modified 2026-06-17.