CVE-2020-16224: Philips Patient Information Center Ix
Medium severity, CVSS 6.5. EPSS: 0.6% probability of exploitation activity in the next 30 days.
In Patient Information Center iX (PICiX) Versions C.02, C.03, the software parses a formatted message or structure but does not handle or incorrectly handles a length field that is inconsistent with the actual length of the associated data, causing the application on the surveillance station to restart.
What to do
No fixed version is stated in NVD's or the CNA's structured data. Check the vendor's advisory among the references.
Affected products
- Philips Patient Information Center Ix: version c.02 only; version c.03 only
Published 2020-09-11. Last modified 2026-06-17.