CVE-2019-25717: Draeger Delta Xl Firmware
Medium severity, CVSS 4.3. EPSS: 0.2% chance of exploitation in the next 30 days.
Dräger Infinity Delta, Delta XL, and Kappa patient monitors contain an information disclosure vulnerability that allows unauthenticated network attackers to access log files over a network connection. Attackers can retrieve device internals, location information, and wired network configuration details from the exposed log files.
Affected products
- Draeger Delta Xl Firmware: affected versions not specified
- Draeger Infinity Delta Firmware: affected versions not specified
- Draeger Kappa Firmware: affected versions not specified
Published 2026-06-02. Last modified 2026-07-22.