CVE-2018-18565: Roche Accu-Chek Inform Ii Firmware
Medium severity, CVSS 6.8. EPSS: 0.5% chance of exploitation in the next 30 days.
An issue was discovered in Roche Accu-Chek Inform II Instrument before 03.06.00 (Serial number below 14000) and 04.x before 04.03.00 (Serial Number above 14000), CoaguChek Pro II before 04.03.00, CoaguChek XS Plus before 03.01.06, CoaguChek XS Pro before 03.01.06, cobas h 232 before 03.01.03 (Serial number below KQ0400000 or KS0400000), and cobas h 232 before 04.00.04 (Serial number above KQ0400000 or KS0400000). A vulnerability in the software update mechanism allows authenticated attackers in the adjacent network to overwrite arbitrary files on the system through a crafted update package.
Affected products
- Roche Accu-Chek Inform Ii Firmware: before 03.06.00 (fixed in 03.06.00); from 04.00.00, before 04.03.00 (fixed in 04.03.00)
- Roche Coaguchek Pro Ii Firmware: before 04.03.00 (fixed in 04.03.00)
- Roche Coaguchek Xs Plus Firmware: before 03.01.06 (fixed in 03.01.06)
- Roche Coaguchek Xs Pro Firmware: before 03.01.06 (fixed in 03.01.06)
- Roche Cobas H 232 Firmware: before 03.01.03 (fixed in 03.01.03); from 04.00.00, before 04.00.04 (fixed in 04.00.04)
Published 2018-11-20. Last modified 2026-06-17.