CVE-2015-3953: Pifzer Plum a+3 Infusion System Firmware

Critical severity, CVSS 9.8. EPSS: 2% chance of exploitation in the next 30 days.

Hard-coded accounts may be used to access Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.

Affected products

  • Pifzer Plum a+3 Infusion System Firmware: up to and including 13.6
  • Pifzer Plum A+ Infusion System Firmware: up to and including 13.4
  • Pifzer Symbiq Infusion System Firmware: up to and including 3.13

Published 2019-03-25. Last modified 2026-06-17.