CVE-2015-1012: Pfizer Lifecare Pca Infusion System Firmware

High severity, CVSS 7.5. EPSS: 0.8% chance of exploitation in the next 30 days.

Wireless keys are stored in plain text on version 5 of the Hospira LifeCare PCA Infusion System. According to Hospira, version 3 of the LifeCare PCA Infusion System is not indicated for wireless use, is not shipped with wireless capabilities, and should not be modified to be used in a wireless capacity in a clinical setting. Hospira has developed a new version of the PCS Infusion System, version 7.0 that addresses the identified vulnerabilities. Version 7.0 has Port 20/FTP and Port 23/TELNET closed by default to prevent unauthorized access.

Affected products

  • Pfizer Lifecare Pca Infusion System Firmware: up to and including 5.0

Published 2019-03-25. Last modified 2026-06-17.